CVE-2015-1012

High

Description

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.

SeverityHigh
CVSS7.5
Exploit probability—
ProductPfizer Lifecare Pca Infusion System Firmware
PublishedMar 25, 2019
First seenMar 25, 2019
Weakness typeCWE-312CWE-312

How to fix

Sources list no fixed version or patch link for this vulnerability.

Affected products

ProductAffected versionsFixed in
pfizer lifecare_pca_infusion_system_firmwarev ≤ 5.0—