CVE-2015-3956

Critical

Description

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

SeverityCritical
CVSS9.8
Exploit probability—
ProductPifzer Plum A\+ Infusion System Firmware
PublishedMar 25, 2019
First seenMar 25, 2019
Weakness typeCWE-345CWE-345

How to fix

Sources list no fixed version or patch link for this vulnerability.

Affected products

ProductAffected versionsFixed in
pifzer plum_a\+_infusion_system_firmwarev ≤ 13.4—
pifzer plum_a\+3_infusion_system_firmwarev ≤ 13.6—
pifzer symbiq_infusion_system_firmwarev ≤ 3.13—