CVE-2018-18563

Critical

Description

An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial Number below KQ0400000 or KS0400000) and cobas h 232 before 04.00.04 (Serial Number above KQ0400000 or KS0400000). Improper access control to a service command allows attackers in the adjacent network to execute arbitrary code on the system through a crafted Poct1-A message.

SeverityCritical
CVSS9.6
Exploit probability—
ProductRoche Accu-chek Inform Ii Firmware
PublishedNov 20, 2018
First seenNov 20, 2018

How to fix

Update
  • roche accu-chek_inform_ii_firmware03.06.00, 04.03.00 or later
  • roche cobas_h_232_firmware03.01.03, 04.00.04 or later
  • roche coaguchek_pro_ii_firmware04.03.00 or later
  • roche coaguchek_xs_plus_firmware03.01.06 or later
  • roche coaguchek_xs_pro_firmware03.01.06 or later

Affected products

ProductAffected versionsFixed in
roche accu-chek_inform_ii_firmwarev < 03.06.00, 04.00.00 ≤ v < 04.03.0003.06.00, 04.03.00
roche cobas_h_232_firmwarev < 03.01.03, 04.00.00 ≤ v < 04.00.0403.01.03, 04.00.04
roche coaguchek_pro_ii_firmwarev < 04.03.0004.03.00
roche coaguchek_xs_plus_firmwarev < 03.01.0603.01.06
roche coaguchek_xs_pro_firmwarev < 03.01.0603.01.06