CVE-2018-18565
Description
An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial number below KQ0400000 or KS0400000), and cobas h 232 before 04.00.04 (Serial number above KQ0400000 or KS0400000). A vulnerability in the software update mechanism allows authenticated attackers in the adjacent network to overwrite arbitrary files on the system through a crafted update package.
How to fix
- roche accu-chek_inform_ii_firmware03.06.00, 04.03.00 or later
- roche cobas_h_232_firmware03.01.03, 04.00.04 or later
- roche coaguchek_pro_ii_firmware04.03.00 or later
- roche coaguchek_xs_plus_firmware03.01.06 or later
- roche coaguchek_xs_pro_firmware03.01.06 or later
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| roche accu-chek_inform_ii_firmware | v < 03.06.00, 04.00.00 ≤ v < 04.03.00 | 03.06.00, 04.03.00 |
| roche cobas_h_232_firmware | v < 03.01.03, 04.00.00 ≤ v < 04.00.04 | 03.01.03, 04.00.04 |
| roche coaguchek_pro_ii_firmware | v < 04.03.00 | 04.03.00 |
| roche coaguchek_xs_plus_firmware | v < 03.01.06 | 03.01.06 |
| roche coaguchek_xs_pro_firmware | v < 03.01.06 | 03.01.06 |
Check your version
Choose the product and enter the version you run to see whether it falls in the affected range.
Attack Surface Management