CVE-2018-4845

High

Description

A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). Remote attackers with either local or remote credentialed access to the "Remote View" feature might be able to elevate their privileges, compromising confidentiality, integrity, and availability of the system. No special skills or user interaction are required to perform this attack. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.

SeverityHigh
CVSS8.8
Exploit probability—
ProductSiemens Rapidpoint 400 Firmware
PublishedJun 26, 2018
First seenJun 26, 2018

How to fix

Update
  • siemens rapidlab_1200_firmware3.3 or later

Affected products

ProductAffected versionsFixed in
siemens rapidpoint_400_firmwareNo range specified—
siemens rapidpoint_500_firmwarev ≤ 2.3, 3.0 ≤ v—
siemens rapidlab_1200_firmwarev < 3.33.3

Check your version

Choose the product and enter the version you run to see whether it falls in the affected range.

In the Intelligence 360 plan

Attack Surface Management

See the assets tied to your domains, the technologies they run and known vulnerabilities in one list.

CVE-2018-4845 · Siemens Rapidpoint 400 Firmware · Dark Radar