CVE-2020-15486
Description
An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.
How to fix
Sources list no fixed version or patch link for this vulnerability.
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| drtrust electrocardiogram_pen_firmware | 2.00.08 | — |
Check your version
Choose the product and enter the version you run to see whether it falls in the affected range.
Attack Surface Management
See the assets tied to your domains, the technologies they run and known vulnerabilities in one list.