CVE-2020-25183

High

Description

Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.

SeverityHigh
CVSS8.0
Exploit probability—
ProductMedtronic Mycarelink Smart Model 25000 Firmware
PublishedDec 14, 2020
First seenDec 14, 2020

How to fix

Sources list no fixed version or patch link for this vulnerability.

Affected products

ProductAffected versionsFixed in
medtronic mycarelink_smart_model_25000_firmwareNo range specified—