CVE-2022-42894
High
Description
A vulnerability has been identified in syngo Dynamics (All versions < VA40G HF01). An unauthenticated Server-Side Request Forgery (SSRF) vulnerability was identified in one of the web services exposed on the syngo Dynamics application that could allow for the leaking of NTLM credentials as well as local service enumeration.
SeverityHigh
CVSS7.5
Exploit probability—
ProductSiemens Syngo Dynamics Cardiovascular Imaging And Information System
PublishedNov 17, 2022
First seenNov 17, 2022
Weakness typeCWE-918Server-side request forgery (SSRF)
How to fix
Update
- siemens syngo_dynamics_cardiovascular_imaging_and_information_systemva40g_hf01 or later
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| siemens syngo_dynamics_cardiovascular_imaging_and_information_system | v < va40g_hf01 | va40g_hf01 |
Check your version
Choose the product and enter the version you run to see whether it falls in the affected range.
In the Intelligence 360 plan
Attack Surface Management