Description

An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks

SeverityMedium
CVSS4.9
Exploit probability49. percentile0.6%
ProductBbraun Battery-pack Sp With Wifi Firmware
PublishedMar 13, 2023
First seenMar 13, 2023
Weakness typeCWE-95CWE-95

How to fix

Sources list no fixed version or patch link for this vulnerability.

Affected products

ProductAffected versionsFixed in
bbraun battery-pack_sp_with_wifi_firmwarev ≤ 053l000092, v ≤ 054u000092—

Check your version

Choose the product and enter the version you run to see whether it falls in the affected range.