CVE-2023-25931

Medium

Description

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.

SeverityMedium
CVSS6.4
Exploit probability—
ProductMedtronic Interstim X Clinician
PublishedMar 1, 2023
First seenMar 1, 2023
Weakness typeCWE-620CWE-620

How to fix

Sources list no fixed version or patch link for this vulnerability.

Affected products

ProductAffected versionsFixed in
medtronic interstim_x_cliniciana51300—
medtronic micro_cliniciana51200—

Check your version

Choose the product and enter the version you run to see whether it falls in the affected range.

In the Intelligence 360 plan

Attack Surface Management

See the assets tied to your domains, the technologies they run and known vulnerabilities in one list.