CVE-2024-9832

Critical

Description

There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

SeverityCritical
CVSS9.3
Exploit probability—
ProductLife2000 Ventilation System
PublishedNov 14, 2024
First seenNov 14, 2024
Weakness typeCWE-307CWE-307

How to fix

Sources list no fixed version or patch link for this vulnerability.

Affected products

ProductAffected versionsFixed in
Baxter Life2000 Ventilation System06.08.00.00 and prior—

Check your version

Choose the product and enter the version you run to see whether it falls in the affected range.

In the Intelligence 360 plan

Attack Surface Management