CVE-2025-0683

Medium

Description

In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.

SeverityMedium
CVSS5.9
Exploit probability—
ProductCMS8000 Patient Monitor
PublishedJan 30, 2025
First seenJan 30, 2025
Weakness typeCWE-359CWE-359

How to fix

Sources list no fixed version or patch link for this vulnerability.

Affected products

ProductAffected versionsFixed in
Contec Health CMS8000 Patient MonitorNo range specified—

Check your version

Choose the product and enter the version you run to see whether it falls in the affected range.

In the Intelligence 360 plan

Attack Surface Management

See the assets tied to your domains, the technologies they run and known vulnerabilities in one list.