CVE-2025-1204

High

Description

The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.

SeverityHigh
CVSS7.7
Exploit probability—
ProductCMS8000 Patient Monitor
PublishedFeb 25, 2025
First seenFeb 25, 2025
Weakness typeCWE-912CWE-912

How to fix

Sources list no fixed version or patch link for this vulnerability.

Affected products

ProductAffected versionsFixed in
Contec Health CMS8000 Patient Monitor0—

Check your version