CVE-2025-14095

Medium

Description

A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software.  Other related CVE's are CVE-2025-14096 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required configuration for Exposure: Physical access to the analyzer is needed. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication.                                                                                                                                                                                        Note: CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.

SeverityMedium
CVSS6.8
Exploit probability—
ProductABL90 FLEX and ABL90 FLEX PLUS Analyzers
PublishedDec 17, 2025
First seenDec 17, 2025

How to fix

Sources list no fixed version or patch link for this vulnerability.

Affected products

ProductAffected versionsFixed in
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS AnalyzersAll application software versions with Windows 7, Windows XP as underlying OS, Application software versions < 3.5MR11 with Windows 10 as underlying OS—
Radiometer Medical Aps AQT90 FLEX AnalyzersAll Application software versions <= 8.13 MR2—
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX AnalyzersApplication software versions < 6.20 MR2 with Windows 7, Windows XP as underlying OS, Application software versions < 6.20 MR2 with Windows 10 as underlying OS—
Radiometer Medical Aps ABL9 AnalyzersApplication software versions < 1.5.0—