CVE-2025-54304

Critical

Description

An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When they are powered on, an X11 display server is started. The display server listens on all network interfaces and is accessible over port 6000. The X11 access control list, by default, allows connections from 127.0.0.1 and 192.168.2.15. If a device is powered on and later connected to a network with DHCP, the device may not be assigned the 192.168.2.15 IP address, leaving the display server accessible by other devices on the network. The exposed X11 display server can then be used to gain root privileges and the ability to execute code remotely by interacting with matchbox-desktop and spawning a terminal. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.

SeverityCritical
CVSS9.8
Exploit probability—
ProductThermofisher Ion Torrent Onetouch 2 Firmware
PublishedDec 4, 2025
First seenDec 4, 2025

How to fix

Affected products

ProductAffected versionsFixed in
thermofisher ion_torrent_onetouch_2_firmwareNo range specified—

Check your version

Choose the product and enter the version you run to see whether it falls in the affected range.

In the Intelligence 360 plan

Attack Surface Management

See the assets tied to your domains, the technologies they run and known vulnerabilities in one list.